Abstract
The paper describes a high precision gas meter suitable for low rates of flow (< 5 cm3 min-1) which can be interfaced with data acquisition software for continuous on-line measurement. The meter used the highly sensitive pressure transducers currently available, operated with a very small back pressure, and had an accuracy around 0.5%. A frequency to voltage converter was developed and built into the gas meter. This allowed direct connection to a data logger/computer without prior computation.
| Original language | English |
|---|---|
| Pages (from-to) | 977-979 |
| Number of pages | 3 |
| Journal | Water Research |
| Volume | 29 |
| Issue number | 3 |
| DOIs | |
| Publication status | Published - Mar 1995 |
Keywords
- high sensitivity gas meter
- low gas flow
- pressure transducer